The Signet 2764-2767 Differential pH & ORP electrodes
are built with the DryLocT connector a Ryton® body and TeflonT
reference junction to handle even the most extreme and
harshest of chemical applications. These differential
electrodes use the field-proven 3-electrode differential
technique: the pH and reference electrodes are measured
against a ground electrode, insuring a steady and stable
signal. The key feature of the differential design is the
reference electrode. The reference element is protected in a
glass half-cell embedded in the reference chamber and is
protected from compounds that may contain sulfides (H2S) and
metals, including lead, mercury, and silver. To ensure a long
service life, the reference features a refillable electrolyte
chamber and a replaceable equitransferant salt bridge, both
easily serviced in the field. The patented porous TeflonT
reference junction resists fouling, clogging and chemical
attack. The other elements of the design are the solution
ground, the pH/ORP electrodes, and the temperature element.
The solution ground eliminates noisy measurements by draining
electrical current away from the reference electrode. The
pH/ORP electrodes are designed with a flat or bulb measurement
surface, and a temperature device that is positioned at the
tip of the measurement surface, making the temperature
response of T95% less than 1 minute. Various temperature
devices offered include 3KO, 300O, or PT-1000 RTD. The
electrodes are used with the Signet 2750 Sensor Electronics,
which provides a blind 4 to 20 mA output or using the digital
(S3L) output may be used to connect to the Signet 8900
instrument. The electrodes can also be used with the Model
2760 preamplifier to connect to the Signet 5700, 8750, or
other suppliers' instruments.
|